| Negotiable | 1 Piece (MOQ) |
| Model | Open System Panall 8000 |
| Sample Throughput | 1~8 samples at the same time |
| Pipetting Range | 20μL ~250μL |
| Detection time | 1~2 hours,relying on the reagent |
| Channel and Available Fluorescein | Channel 1: FAM, SYBR Green I, etc. Channel 2: VIC, HEX, TET, JOE, etc. Channel 3: ROX, Texas Red, etc. Channel 4: Cy5, etc. |
| Pipetting Performance | 20μL ≤ V<40μL: accuracy: A≤5.0%, repeatability: CV≤3.0% 40μL ≤ V<100μL: accuracy: A≤3.0%, repeatability: CV≤1.5% V ≥100μL: accuracy: A≤1.0%, repeatability: CV≤1.0% |
| Extraction Heating Rate | Average heating rate: ≥ 1.5ºC/s; |
| Extraction Temperature Accuracy | ≤ 1.0ºC |
| PCR Heating Rate | Average heating rate: ≥ 4.5ºC/s; Maximum heating rate: ≥ 6.1ºC/s |
| PCR Cooling Rate | Average cooling rate: ≥ 3.5ºC/s; Maximum cooling rate: ≥ 5.0ºC/s. |
| PCR Temperature Accuracy | ≤ 0.1ºC |
| Sample Testing Linearity and Repeatability | Linear correlation: /r/ ≥ 0.998 Repeatability: CV≤ 1.5% |
| Information Management | Sample information: with scanner inside, Panall 8000 can scan and record the sample information; Reagent information: visual system can automatically identify the kit information and run the corresponding program |
| Minimized Contamination | Directional exhaust & negative pressure system; HEPA filtration; UV disinfection; Shortest fixed path for sample operation |
| Data Storage | 1000 experiment files can be stored. |
| Language | Chinese and English |
| Power Supply and Power Consumption | AC 100~240V, 50/60Hz; 1200VA |
| Communication Specification | Internet Port: TCP/IP protocol, Ethernet connection USB Port: USB 2.0 |
| Dimension | 750mm(L) × 350mm (W) × 600mm (H) |
| Weight | 80Kg |